Determination of response of real-time SidePak AM510 monitor to secondhand smoke, other common indoor aerosols, and outdoor aerosol

J Environ Monit. 2011 Jun;13(6):1695-702. doi: 10.1039/c0em00732c. Epub 2011 May 17.

Abstract

The amount of light scattered by airborne particles inside an aerosol photometer will vary not only with the mass concentration, but also with particle properties such as size, shape, and composition. This study conducted controlled experiments to compare the measurements of a real-time photometer, the SidePak AM510 monitor (SidePak), with gravimetric mass. PM sources tested were outdoor aerosols, and four indoor combustion sources: cigarettes, incense, wood chips, and toasting bread. The calibration factor for rescaling the SidePak measurements to agree with gravimetric mass was similar for the cigarette and incense sources, but different for burning wood chips and toasting bread. The calibration factors for ambient urban aerosols differed substantially from day to day, due to variations in the sources and composition of outdoor PM. A field evaluation inside a casino with active smokers yielded calibration factors consistent with those obtained in the controlled experiments with cigarette smoke.

Publication types

  • Evaluation Study

MeSH terms

  • Aerosols / analysis*
  • Air Pollutants / analysis*
  • Air Pollution, Indoor / analysis*
  • Environmental Monitoring / instrumentation*
  • Particulate Matter / analysis
  • Tobacco Smoke Pollution / analysis*

Substances

  • Aerosols
  • Air Pollutants
  • Particulate Matter
  • Tobacco Smoke Pollution